Abstract
動機背景:隨著時代的演進,實驗室為了提高操作效率、錯誤檢測、患者安全及提升檢驗時效(turnaround time, TAT)等因素開始建置自動驗證 (Autoverification, AV, automated result verification)系統。利用實驗室資訊系統 (laboratory information system, LIS)系統或透過中介軟體(Middleware)來控制檢驗報告是否自動發送。依據台灣醫檢學會(Taiwan Society of Laboratory Medicine, TSLM)所建立的急診(室)檢驗(emergency test)時效,其定義為收到檢體至完成報告時間為30分鐘。在健保給付上,急診(室)檢驗如能在開立檢驗後1小時內完成報告,將有健保額外加成20%,這對於醫院將是可觀的收入。急診生化檢驗依某醫院體系(案例醫院之體系)間訂立標準為電腦(LIS)檢體收件至報告完成(TAT)30分鐘達成率指標值為 ≧90%,台灣中部某區域醫院(案例醫院)急診室生化於檢驗醫學部無法穩定達到檢驗時效指標,故想藉由透過資訊端導入自動化驗證系統藉此改善急診生化檢驗時效,並提高健保給付。研究材料與方法:以中部某區域醫院(案例醫院)導入自動化驗證(AV)系統為例,首先分析導入AV系統的工作流程,先向其他有導入AV系統的醫院(尤其是總院),學習方向並參考其經驗,了解目前單位內所面臨到的問題,其次博覽各種國、內外文獻及各種指導方針(guidelines),如美國病理學會(College of American Pathologists, CAP)、臨床與實驗室標準協會(Clinical & Laboratory Standards Institute, CLSI)、台灣醫檢學會等作為指引。AV的設計依Delta Check(A)、Auto extended range(U)、檢體種類(T)、儀器異常註記(O)、檢體狀態(E)等為數據過濾器(組成要件),最後經由各種AV的驗證完成系統的建置與正式啟用。結果與討論:案例醫院,利用急診(室)生化總醫令數進行加權後,經由獨立樣本t檢定,結果發現在導入AV前5個月TAT 30分鐘達成率為89.69%,報告完成平均時間(TAT)為17.79分鐘,導入AV後5個月達成率平均顯著提升為91.85%(P <0.001),TAT顯著降為16.78分鐘(P <0.001),達成率顯著提高而TAT顯著下降。急件生化(Stat Chemistry,包含門診、急診、病房等被要求急作的生化,共27項生化項目)啟用AV 5個月約省下2370分鐘(約1706元),2020年若全年實施AV,可節省約5886分鐘(約21190元)。若全部生化(急及非急件生化)皆導入,可節省約19022分鐘(約68472元)。利用投資報酬率(return on investment, ROI)來計算實驗室AV導入之後的總效益。全部生化投資報酬率為13.20%,如單以急件生化來說,ROI為2.84%。AV自動發送報告必須是Delta Check(A)通過、Auto extended range(U)通過、檢體種類(T)為血液、儀器註/標記(O)無出現、檢體狀態(E)通過(非科收狀態、檢體不足、凝固、醫令刪除或其中一項儀器備註欄位有儀器flag,導致整筆同安全身分標識(Security identifiers, SID)之生化項目被攔截)。急件生化啟用5個月,不通過的部分A占14%(30939件)、U佔18%(39951件)、O佔3.3%(7203件)、E佔17.2%(38097件)、T佔0%(0件),全部通過佔64.2%,即AV pass rate 64.2%,優於大部分文獻紀載。AV導入後5個月TAT顯著降為16.78分鐘(P <0.001),約可節省1.01分鐘(5.67%),與一般文獻相比,進步幅度偏低。原因可能為台灣有TSLM指標TAT 30分鐘,原TAT就已相對較短,因此改善幅度較小。
Background: As the times change, clinical laboratories have begun to build automated result verification (Autoverification, AV) systems in order to improve work efficiency, error detection, patient safety, and improve turnaround time (TAT). Use LIS (laboratory information system) system or through middleware to control whether the medical examination reports is verified automatically. According to laboratory report TAT in the emergency test by the Taiwan Society of Laboratory Medicine(TSLM), it is defined as the time from receipt of the specimen to completion of the report is 30 minutes. In terms of National Health Insurance(NHI) payment, if the emergency medical examination can complete the report within one hour after the examination is issued, there will be a NHI premium of 20%, which will be a considerable income for the hospital. According to the standard established by the certain case hospital system, the 30-minute target value ratio of the emergency chemical test from the specimen computer sign to the report completed is ≧90%. The emergency chemical test of the case Regional Hospital in central Taiwan(the case hospital) cannot stably achieve TAT index, so I want to build an AV system to improve the TAT of emergency chemical test and increase NHI payments.Material and Method: Taking the introduction of an AV system into the case regional hospital in central Taiwan as an example(the case hospital), first analyze the workflow of importing the AV system, and go to other hospitals (especially the general hospital) that have leaded-in the AV system as a learning direction to understand the current laboratory problems encountered internally. Secondly, widely read various domestic, foreign documents and various guidelines, such as CAP, CLIA, TSLM, etc. The AV design is based on delta check (A), auto extended range (U), specimen type (T), instrument abnormality flag (O), specimen status (E), etc. as data filters, and finally completed by various AV verifications the establishment and official launch of the system.Result and Discussion: In the case hospital, after weighting the total number of emergency chemical medical orders, through independent sample t-test, it was found that the TAT 30-minute target value ratio was 89.69% during 5 months before AV activation, and TAT was 17.79 minutes. During five months after the introduction of AV, the TAT target value ratio increased significantly to 91.85% on average (P <0.001) and TAT significantly reduced to 16.78 minutes(P <0.001). The TAT target value ratio increased significantly and TAT decreased significantly.Emergency chemistry will save about 2370 minutes (approximately 1706 NT$) during 5 months by using AV. If AV is implemented throughout the year in 2020, it will save approximately 5886 minutes (approximately 21,190 NT$). If all chemical tests are imported AV (27 items), about 19022 minutes (about 68472 NT$) can be saved.Use ROI(return on investment) to calculate the total benefits of the laboratory after starting to use AV. The ROI of all chemical tests was 13.20%. For the Stat chemical tests, the ROI is 2.84%.The AV automatic transmission report must be delta check (A) passed, auto extended range (U) passed, specimen type (T) is blood, instrument’s flag (O) is not present, specimen status (E) passed (Non-computer signing status, insufficient sample volume, clot, medical order deletion, or one of the same SID tests had an instrument’s flag that led to the block of the whole stroke ). It has been 5 months since the AV system was launched in the stat chemical tests, AV system components fail 14% of A (30,939 tests), 18% of U (39951 tests), 3.3% of O (7203 tests), 17.2% of E (38097 tests), and 0% of T (0 tests). All the components passed was 64.2%.The AV pass rate of 64.2% is better than most documents.During 5 months after AV activation, TAT was significantly reduced to 16.78 minutes(P <0.001). Approximately 1.01 minutes (5.67%) can be saved, which is relatively low compared with the general literature. The reason may be that Taiwan has a TAT index of 30 minutes, and the original TAT is already relatively short, so the improvement is small.
Subjects
自動驗證
檢驗時效
TAT 30分鐘達成率
急診室生化檢驗
急件生化
Autoverification
automated result verification
turnaround time
TAT 30-minute target value ratio
emergency chemical test
Stat chemistry
Type
thesis
File(s)
No Thumbnail Available
Name
109CSMU5743011-001.zip
Size
1.37 MB
Format
Unknown
Checksum
(MD5):9ce0d82e2530d562a527e77fa7ab2118