使用動態田口方法探討最佳化TLD 熱發光 計讀系統應用乳癌近接治療之劑量評估

Date Issued
2024
Author(s)
宋瑜庭
SUNG, YU-TING
Advisor
陳健懿
CHEN, CHIEN-YI  
URI
https://handle.ncl.edu.tw/11296/655jqh
https://scholars.csmu.edu.tw/handle/123456789/7558
Abstract
本研究利用動態田口方法優化熱發光劑量計(TLD)系統,用於乳癌192Ir近接治療的劑量評估。為確保治療的精準性和安全性,我們收集了TLD-100H量測系統的相關資料,並使用動態田口方法和L18直交表篩選出最佳計讀條件組合:初始溫度130°C、加熱速率3°C/sec、預熱時間6秒、最高溫度220°C及持續26秒。根據校正曲線方程式y(cGy/treatment)=0.7398x(μC)+0.204,R平方值0.9959,可以非常準確地將電量轉換為劑量值。在乳癌近接治療的劑量量測中,假體乳房橫軸皮膚劑量隨距離增加逐漸降低。在距離治療中心1.5公分處設置27個測量點,每點含三顆TLD。測量發現,距離治療中心1.5公分的A點劑量值為11.9 ± 1.78 Gy,距離6公分的D點劑量降至3.00 ± 0.45 Gy,距離13.5公分的E點劑量進一步降至0.57 ± 0.09 Gy。同樣,在R方向上,距離12公分處的B點劑量為0.87 ± 0.13 Gy,而19.5公分處的C點劑量降至0.31 ± 0.05 Gy。本研究的最佳化計讀條件組合顯著提高了TLD-100H的靈敏度和再現性,確保了準確的劑量測量。這對乳癌近接治療的劑量評估具有重要意義,幫助臨床醫生確定適當治療劑量,提高治療效果,同時減少患者的不良影響。
This study employs the dynamic Taguchi method to optimize the thermoluminescent dosimeter (TLD) system for dose assessment in breast cancer 192Ir brachytherapy. Breast cancer is the most prevalent cancer among women worldwide, and brachytherapy delivers high radiation doses directly to the tumor while minimizing damage to surrounding tissues. To ensure precision and safety in treatment, we gathered data on the TLD-100H measurement system and applied the dynamic Taguchi method along with an L18 orthogonal array to identify optimal reading conditions: an initial temperature of 130°C, heating rate of 3°C/sec, preheat time of 6 seconds, maximum temperature of 220°C, and duration of 26 seconds. Using the calibration curve equation y(cGy/treatment) = 0.7398x(μC) + 0.204, with an R-squared value of 0.9959, dose values can be accurately converted from charge measurements.In the dose measurement for breast cancer brachytherapy, the skin dose on the transverse axis of the breast phantom decreased with increasing distance from the treatment center. We placed 27 measurement points, each containing three TLDs, 1.5 cm from the treatment center. Results indicated that the dose at point A, 1.5 cm from the center, was 11.9 ± 1.78 Gy. At point D, 6 cm away, the dose decreased to 3.00 ± 0.45 Gy, and at point E, 13.5 cm away, it further decreased to 0.57 ± 0.09 Gy. Similarly, in the R direction, the dose at point B, 12 cm away, was 0.87 ± 0.13 Gy, and at point C, 19.5 cm away, it dropped to 0.31 ± 0.05 Gy.The optimized reading conditions significantly enhanced the sensitivity and reproducibility of the TLD-100H, ensuring precise dose measurement. This is crucial for dose assessment in breast cancer brachytherapy, assisting clinicians in determining appropriate treatment doses, improving therapeutic effectiveness, and minimizing adverse effects on patients.目錄
Subjects
熱釋光劑量計
動態田口方法
乳癌
近接治療
侖道假體
銥192
TLD-100H
Taguchi Methods
Breast cancer
Brachytherapy
Rando phantom
Ir-192
Type
thesis
File(s)
No Thumbnail Available
Name

112CSMU0770005-002.pdf

Size

10.79 MB

Format

Adobe PDF

Checksum

(MD5):9d7c4c15e36614ff423bac55401f444b

Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your Institution's web identity.

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback