Evaluating air kerma at a medical cyclotron after refined shielding design using the TLD approach
Journal
Journal of Nuclear Science and Technology
Date Issued
2013-06-01
Author(s)
Abstract
Cyclotrons are increasingly constructed at medical universities in Taiwan to produce nuclear medicine. However, their operation generates amounts of highly energetic photons from 18O(p, γn)18F reactions. This reaction posed a health hazard to the public. Chung Shan University Hospital had to increase its 2-deoxy-2-[18F]-fluoro-D-glucose products in 2011 to 11.5 times that in 2004 and redesign the self-shielded (CTI) Radioisotope Delivery System (RDS)-111 to meet guild requirements regarding the production of positron emission tomography (PET). This study is the first to evaluate air kerma in a CTI RDS-111 cyclotron center with a thermoluminescent dosimeter (TLD). Air kerma of up to 17.0±2.8 mGy/mo indicated that heavy leakage of photons penetrated the PET/CT facility. The new L-shaped concrete shield and added d2 door can adequately protect the public. The minimum detectable limit of air kerma is discussed to demonstrate the reliability of the TLD approach.