Abstract
Elevated urinary cystine levels are closely associated with the development of cystine stone. Therefore, the ability to rapidly and efficiently determine urinary cystine levels is crucial for physicians to manage patients with cystinuria or those undergoing cystine medication. In this study, an amperometric method employing a commercial screen-printed silver electrode was successfully established. The resulting calibration curve indicated a detection limit of 0.65 mg/dL. Satisfactory recoveries ranging from 89% to 109% were obtained for urine samples. The method was also effective for the quality control analysis of cystine in pharmaceutical tablets. The recovery of cystine from pharmaceutical tablets ranged from 98% to 101% using the developed method. This method enables the rapid and accurate determination of cystine in both urine samples and pharmaceutical tablets and provides valuable information for clinical diagnosis and pharmaceutical quality control.