救護車負壓排氣效能評估-以改裝市售系統為例

Date Issued
2026
Author(s)
姚辰龍
YAO,CHEN-LONG
Advisor
賴全裕
LAI,CHANE-YU  
URI
https://handle.ncl.edu.tw/11296/t8j72m
https://scholars.csmu.edu.tw/handle/123456789/6446
Abstract
隨著 SARS、MERS 及 COVID-19 等空氣傳染性疾病持續出現,緊急醫療服務(Emergency Medical Services, EMS)於到院前醫療運輸階段之感染控制風險日益受到重視。救護車艙體空間狹小、通風受限,且救護人員需近距離執行醫療處置,使其成為生物性氣膠高度暴露之高風險環境。現行多數救護車仍主要仰賴個人防護裝備與一般通風系統,缺乏能於污染源端即時控制氣膠擴散之工程控制措施。此一防護模式高度仰賴人員正確穿戴與操作行為,且於長時間勤務、高溫密閉或高負荷醫療處置情境下,易增加防護失效與人員職業暴露風險,顯示僅以個人防護裝備作為主要防疫手段仍存在實務與結構性限制。本研究以職業衛生層級控制原則為基礎,評估一套改良型可攜式負壓抽吸系統於救護車艙內之生物性氣膠控制效能。研究以市售 3M™ AIR-MATE™ HEPA 過濾系統為核心,進行風扇、流量控制與氣罩結構之工程改良,使其可於病患近端形成穩定之定向抽吸氣流。實驗採用枯草桿菌(Bacillus subtilis)作為生物氣膠指標,透過霧化裝置模擬病患呼吸與咳嗽產生之氣膠,並於實際救護車艙內進行噴菌測試。本研究之設計概念並非建構整體救護車艙體負壓環境,而是以近源抽吸為核心,於不變更既有車體結構之前提下,降低氣膠向救護人員呼吸區擴散之風險,兼顧實務可行性與部署彈性。空氣中微生物濃度以安德森六階生物氣膠採樣器量測,分析不同氣罩配置(無氣罩、雙側包圍氣罩、全包圍氣罩)及操作條件下之菌落形成單位(CFU/m³)與濃度衰減量。無氣罩情境下雖呈現較高之濃度,顯示氣膠已大量擴散至救護人員呼吸區。相對地,雙側包圍氣罩、全包圍氣罩配置可大幅降低採樣點之菌落數,證實近源負壓抽吸能有效限制氣膠逸散,降低人員暴露風險。枯草桿菌具備培養穩定、可操作性高及符合生物安全規範等特性,常被應用於生物氣膠工程控制效能評估,能合理反映氣膠於密閉空間中之傳播與沉降行為。綜合而言,本研究證實改良型負壓抽吸系統結合氣罩設計,能作為救護車艙內感染控制之有效工程措施,並突顯僅以濃度衰減指標評估防護效果之限制。研究成果可作為未來救護車感染控制設備設計、效能驗證及政策制定之重要參考。研究進一步顯示,救護車感染控制若僅以「車艙整體換氣次數(ACH)」或單一通風指標作為防護效能判斷依據,仍存在實務盲點。由於救護車艙體空間狹小且醫療處置高度集中於病患近端,氣膠暴露風險具有明顯之「近源性」與「時間敏感性」特徵。相較於全面改裝車體形成整艙負壓環境,本研究所採用之近源抽吸策略,能在不變更既有車體結構之前提下,即時攔截病患呼吸或咳嗽所產生之生物性氣膠,降低其向救護人員呼吸區擴散的機率,具備高度實務可行性。 研究結果顯示,在未設置氣罩之情境下,氣膠可迅速於救護車艙內擴散,導致救護人員呼吸區測得之菌落濃度維持於高水平,反映出僅依賴一般通風或車艙換氣,難以有效控制近距離暴露風險。相對而言,當負壓抽吸系統結合雙側包圍式或全包圍式氣罩設計時,於固定作用時間內可觀察到明顯之濃度衰減,顯示氣膠於逸散初期即被有效捕集。此結果支持工程控制應優先部署於污染源端,而非僅作為整體環境通風之輔助設施。 此外,透過風量旋鈕與實測風量之線性校正,驗證改良後系統於救護車實際供電條件下,仍可維持穩定輸出,確保不同操作階段具備可重現之抽吸效能。噪音測試結果亦顯示,於有效捕集氣膠之操作風量範圍內,系統噪音仍可控制於臨床可接受水準,不致干擾醫療處置或人員溝通。上述結果顯示,本研究所提出之改裝型可攜式負壓抽吸系統,不僅具備工程控制效能,亦兼顧救護實務中對於機動性、穩定性與人因操作之需求。 綜合而言,本系統不僅驗證近源負壓抽吸於救護車艙內感染控制之可行性,亦建立一套可供後續設備開發與效能驗證參考之測試流程與指標,對於提升到院前醫療運輸之職業安全與感染防護品質,具有實質貢獻。
With the continued emergence of airborne infectious diseases such as SARS, MERS, and COVID-19, infection control risks during the prehospital medical transport phase of emergency medical services have received increasing attention. ambulance compartments are characterized by limited space and restricted ventilation, and emergency medical personnel are required to perform medical procedures at close proximity to patients, making these environments high-risk settings for bioaerosol exposure. most ambulances currently rely primarily on personal protective equipment and general ventilation systems, while lacking engineering control measures capable of controlling aerosol dispersion at the source. this protection strategy is highly dependent on correct use and operation by personnel, and under conditions such as prolonged duty periods, high-temperature enclosed environments, or high-intensity medical procedures, the risk of protection failure and occupational exposure may increase, indicating practical and structural limitations of relying solely on personal protective equipment as the primary infection prevention measure.based on the hierarchy of controls in occupational hygiene, this study evaluated the bioaerosol control performance of a modified portable negative-pressure suction system within an ambulance compartment. a commercially available 3M™ AIR-MATE™ HEPA filtration system was used as the core unit, and engineering modifications were applied to the fan, airflow control, and hood structure to enable the formation of a stable, directional suction airflow near the patient. Bacillus subtilis was selected as the bioaerosol indicator, and aerosols simulating patient breathing and coughing were generated using a nebulization device. bioaerosol release experiments were conducted inside an actual ambulance compartment. rather than constructing a whole-compartment negative-pressure environment, the system design focused on near-source suction to reduce the dispersion of aerosols toward the breathing zone of emergency medical personnel without altering the existing vehicle structure, thereby maintaining practical feasibility and deployment flexibility.airborne microbial concentrations were measured using a six-stage Andersen bioaerosol sampler, and colony-forming units per cubic meter and concentration decay were analyzed under different hood configurations, including no hood, a dual-sided enclosure hood, and a fully enclosed hood. the results showed that under the no-hood condition, higher concentrations were observed, indicating extensive aerosol dispersion into the breathing zone of emergency medical personnel. in contrast, both the dual-sided enclosure hood and fully enclosed hood configurations substantially reduced colony counts at the sampling location, confirming that near-source negative-pressure suction effectively limited aerosol dispersion and reduced personnel exposure. Bacillus subtilis exhibits stable cultivation characteristics, high operability, and compliance with biosafety regulations, and is commonly used in engineering evaluations of bioaerosol control performance, reasonably representing aerosol transport and deposition behavior in confined environments.overall, this study demonstrates that a modified negative-pressure suction system combined with hood design can serve as an effective engineering control measure for infection control within ambulance compartments, while also highlighting the limitations of evaluating protective performance solely based on concentration decay metrics. the findings further indicate that if ambulance infection control relies only on whole-compartment air changes per hour or a single ventilation indicator as the basis for performance assessment, practical blind spots may remain. due to the confined space of ambulance compartments and the concentration of medical procedures near the patient, aerosol exposure risk exhibits pronounced source proximity and time sensitivity. compared with comprehensive vehicle modification to create a whole-compartment negative-pressure environment, the near-source suction strategy adopted in this study enables immediate interception of bioaerosols generated by patient breathing or coughing without altering the existing vehicle structure, demonstrating high practical applicability.in addition, linear calibration between the airflow control dial and measured airflow verified that the modified system could maintain stable output under actual ambulance power supply conditions, ensuring reproducible suction performance across different operational stages. noise measurements also showed that within the airflow range effective for aerosol capture, system noise levels remained within clinically acceptable limits and did not interfere with medical procedures or personnel communication. these results indicate that the proposed modified portable negative-pressure suction system provides effective engineering control while also addressing mobility, operational stability, and human factors requirements in prehospital emergency care.collectively, this system not only verifies the feasibility of near-source negative-pressure suction for infection control within ambulance compartments but also establishes a testing process and performance indicators that can serve as references for subsequent equipment development and efficacy evaluation. the findings contribute substantially to improving occupational safety and infection prevention quality in prehospital medical transport.
Subjects
HEPA
感染控制
生物氣膠
工程控制
負壓抽吸
救護車
過濾
infection control
bioaerosol
engineering control
negative-pressure suction
ambulance
filtration
Type
thesis
File(s)
No Thumbnail Available
Name

113CSMU074F006-001.pdf

Size

5.43 MB

Format

Adobe PDF

Checksum

(MD5):64633fb4eaa83efe1323627942f9cfac

Make it your own

DSpace-CRIS can be extensively configured to meet your needs. Decide which information need to be collected and available with fine-grained security. Start updating the theme to match your Institution's web identity.

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Cookie settings
  • Privacy policy
  • End User Agreement
  • Send Feedback