Options
The Accuracy of Ultrasonic Estimation of Bladder Volume
Other Title
A Comparison of Portable and Stationary Equipment
Journal
Archives of Physical Medicine and Rehabilitation
Date Issued
2004-01-01
Author(s)
Abstract
Objective: To compare the accuracy of portable and stationary ultrasonography equipment in estimating residual bladder volume. Design: A prospective study. Setting: A rehabilitation hospital affiliated with a medical university in Taiwan. Participants: Sixty-four patients with voiding dysfunction. Interventions: Each patient received both types of ultrasonography, plus catheterization (as the criterion standard for measuring bladder volume). Main Outcome Measures: The mean errors and mean percentage error resulting from use of the 2 types of equipment were recorded. Results: Stationary real-time ultrasonography had a lower absolute error (21.9±25.0mL vs 34.4±38.2mL) and percentage error (13.1%±10.5% vs 36%±52.4%) than the portable ultrasonography (P<.05). Conclusion: Stationary ultrasonography provided a more accurate estimate of residual bladder volumes than did the portable device. However, portable ultrasonography is more convenient to use and has an acceptable accuracy for clinical use.
Type
journal article