丁健益Ting, Chien-Yi羅子玲Lo, Tzu-LingTzu-LingLo醫學影像暨放射科學系碩士班2026-03-262029-09-0120252024-07-02https://handle.ncl.edu.tw/11296/86scvbhttps://scholars.csmu.edu.tw/handle/123456789/7565隨著電腦斷層掃描 (CT) 的發展,需求逐漸增加,對人體造成的輻射劑量一直受到高度關注,尤其是對於嬰幼兒這類特殊族群,對輻射影響的關注日益增加。嬰幼兒與成人在解剖與生理結構上的差異,導致其對輻射的敏感度不同。嬰幼兒的先天性疾病和流感等廣泛性感染,有時需要進行 CT 掃描作為進一步診斷工具,而不僅依賴常規X光檢查。本研究探討了 CT掃描對新生兒器官的吸收劑量 和全身有效劑量,使用 Sun Nuclear 公司生產的 CIRS ATOM Newborn Model 703 擬人假體模型,結合熱發光劑量計 (TLD) 進行實測,並與劑量運算軟體 VirtualDose進行比較,評估兩者間的劑量差異。根據 ICRP第103 號組織加權因數的計算結果顯示,未屏蔽時得出的全身有效劑量為 0.8692 mSv (0.30±0.24 mGy);屏蔽後為 0.8368mSv (0.28±0.27mGy),劑量減少了 4%,與 VirtualDose相比,各器官的變異度範圍從 10%到125%不等。因 ICRP 第 60 號與 103 號報告中的組織加權因數的不同,我們重新計算了 AAPM 第 96 號報告中的 k 因子轉換係數,根據實驗結果與其他學者相比,差異僅 18%。這份研究模擬了臨床參數與新生兒身體組織的結構和特徵,以及可靠的劑量評估方式,有助於確保新生兒曝露於輻射時的安全性。As computed tomography (CT) technology has advanced, its demand has increased, and the radiation dose to the human body, particularly for infants, has become a significant concern. The anatomical and physiological differences between infants and adults result in varying sensitivities to radiation. Infants with congenital diseases or widespread infections like influenza often require CT scans for further diagnosis, beyond standard Xrays. This study examines the absorbed dose and whole-body effective dose from CT scans on newborns using the Sun Nuclear CIRS ATOM Newborn Model 703 phantom and thermoluminescent dosimeters (TLD), comparing the results with those from VirtualDose software. Based on tissue weighting factors in ICRP Report No. 103, the unshielded wholebody effective dose was 0.8692 mSv (0.30±0.24 mGy), and with shielding, it was 0.8368 mSv (0.28±0.27 mGy), a 4% reduction. The variation between organs compared to VirtualDose ranged from 10% to 125%. Due to differences in tissue weighting factors between ICRP Reports No. 60 and No. 103, we recalculated the k-factor from AAPM Report No. 96, with only an 18% difference compared to previous studies. This study simulated clinical parameters and the structure and characteristics of newborn body tissues, using reliable dose assessment methods. The results contribute to ensuring the safety of newborns exposed to radiation新生兒胸部電腦斷層劑量轉換因子劑量模擬熱發光劑量計NewbornChest CTDose Conversion Factorsk-factorVirtualDoseTLD新生兒胸部電腦斷層輻射劑量與劑量轉換因子之研究thesis