陳賢堂;孫涵瑛Shyan-Tarng Chen;Han-Ying Sun林建年Chien-Nien Lin視光學系2026-03-242025/8/12022https://handle.ncl.edu.tw/11296/hfmnxkhttps://scholars.csmu.edu.tw/handle/123456789/6514目的:新型光場技術結合虛擬實境系統量測屈光度數,與傳統驗光設備檢測數值進行比較,驗證其有效性。研究方法:共招募50位參與者進行實驗,平均年齡為22.12 ± 2.2歲。使用檢影鏡、自動驗光儀(閉鎖式、開放式)、自覺式驗光和光場虛擬實境系統等方式量測屈光度數,將測量度數轉換為等價球面、J0、J45、前主徑線和後主徑線,利用組內相關係數(Intraclass correlation coefficient, ICC)進行一致性與重複性的統計分析。結果:在一致性方面,光場虛擬實境系統與閉鎖式自動驗光儀相比,等價球面的ICC值為85.2% (CI: 0.79 to 0.90);前主徑線為86.1% (CI: 0.80 to 0.90);後主徑線為83.9% (CI: 0.77 to 0.89)。光場虛擬實境系統與開放式自動驗光儀相比,等價球面的ICC值為85.5% (CI: 0.79 to 0.90);前主徑線為86.7% (CI: 0.81 to 0.91);後主徑線為83.8% (CI: 0.77 to 0.89)。光場虛擬實境系統與檢影鏡相比,等價球面的ICC值為82.7% (CI: 0.75 to 0.88);前主徑線為84.3% (CI: 0.77 to 0.89);後主徑線為80.7% (CI: 0.73 to 0.87)。虛擬光場實境系統與自覺式驗光相比,等價球面的ICC值為84.7% (CI: 0.78 to 0.89);前主徑線為86.4% (CI: 0.80 to 0.91);後主徑線為82.0% (CI: 0.74 to 0.88)。在重複性方面,前、後主徑線皆達顯著高度一致性ICC值為97.5% (CI: 0.97 to 0.98)、88.8% (CI: 0.85 to 0.92)。結論:光場虛擬實境系統測量屈光度數的數值與傳統驗光設備皆達高度一致性與重複性。由於採用虛擬實境系統,可以解決空間限制及搬運上的困難,在未來或許可以將此系統列入快速量測屈光度數的初級保健設備。Purpose:To conduct clinical validation of a light field virtual reality technology (LFVR) that is able to evaluate the refractive error.Methods:A total of fifty participants were recruited into the experiment. The average age was 22.12 ± 2.2 years. For the agreement assessment, spherical equivalent (SE), front and back focal line (F2, F1) measurements were evaluated from retinoscopy (RET), closure autorefractor (CA), open-field autorefractor (OFA), subjective refraction (SR) and LFVR, respectively. Results:With regard to agreement, the intraclass correlation coefficient (ICC) value for the SE between the LFVR and CA was 85.2% (CI: 0.79 to 0.90), OFA was 85.5% (CI: 0.79 to 0.90), RET was 82.7% (CI: 0.75 to 0.88), SR was 84.7% (CI: 0.78 to 0.89), respectively. The ICC value for the F1 between the LFVR and CA was 83.9% (CI: 0.77 to 0.89), OFA was 83.8% (CI: 0.77 to 0.89), RET was 80.7% (CI: 0.73 to 0.87), SR was 82.0% (CI: 0.74 to 0.88), respectively. The ICC value for the F2 between the LFVR and CA was 86.1% (CI: 0.80 to 0.90), OFA was 86.7% (CI: 0.81 to 0.91), RET was 84.3% (CI: 0.77 to 0.89), SR was 86.4% (CI: 0.80 to 0.91), respectively. In relation to precision, intra-repeatability for both F1 and F2 were significant high agreement 88.8% (CI: 0.85 to 0.92) and 97.5% (CI: 0.97 to 0.98).Conclusion:LFVR showed strong agreement and repeatability with CA, OFA, RET and SR in the F1 and F2. LFVR could be the primary eye care instrument to measure refractive errors.屈光不正虛擬實境光場技術組內一致性Refractive errorVirtual realityLight fieldIntraclass correlation coefficient使用新型虛擬實境系統測量眼球屈光度thesis