TOU-RONG CHENYeu-Sheng Tyan2026-10-162026-10-162011-12-01Pure ID: 15278296Pure UUID: bb036769-858f-4cb5-aa80-a458fb4c456910.1016/j.radmeas.2011.09.010https://www.scopus.com/pages/publications/82355181058https://scholars.csmu.edu.tw/handle/123456789/30174The purpose of this study is to estimate the dose reduction of bismuth shielding for pediatric CT using a real-time CT dose probe. Routine CT examinations of head, chest and abdomen were performed on a 10-year pediatric anthropomorphic phantom, with and without bismuth shield. A solid-state dosimeter was used to obtain real-time dose measurements and to estimate the dose reduction. The maximum dose reduction of eyes using bismuth shield for in-plane CT scanning can achieve 56% during spiral head scanning interval. The average dose reduction is around 25% for in-plane CT scanning, and reaches 62% combining the technique of automatic tube current modulation. The solid-state probe can provide fast and convenient measurements for dose estimation of superficial organs. It is suggested to use Bismuth shielding in routine CT examinations to protect the radiosensitive tissues.5enReal-time estimation of dose reduction for pediatric CT using bismuth shieldingjournal article